SBIR/STTR Research Award
Pneumatic Focusing Gas Chromotography: A 3-in-1 Continuous, Automated, Ambient-Fenceline-Fugitive Emissions Instrument
Award amount
$225,000Program and phase
SBIR · Phase IIAward year
2005Award details
- Awarding agency
- Environmental Protection Agency
- Branch
- Not reported
- Recipient company
- VOC Technologies
- Proposal award date
- 2005-04-01
- Contract end date
- 2006-06-30
- UEI
- Not reported
- Solicitation number
- PR-NC-04-10484
- Topic code
- 04-NCER-P2
Research abstract
VOC Technologies, Inc. (VOCTEC) developed a patent-pending new technology
for analysis of volatile organic compounds (VOCs) and hazardous air pollutants
(HAPs) by compressing an air sample to high pressure before injecting it into
a specially designed gas chromatograph (GC). This GC is constructed entirely
within a personal computer (the “GC-in-a-PC”), which operates unattended
and indefinitely in the field with network communication. Compression both
concentrates the sample and removes water vapor by condensation. This procedure
is called pneumatic focusing gas chromatography (PFGC).
This highly automated new technology has the potential to lower the cost of
a VOC/HAP analysis by a factor of 100, requiring no special operator skills.
With PFGC, a 300 cc sample can be injected directly on the column and achieve
a 50 ppb level of detection (LOD) for benzene, with 500,000 theoretical plates
using either flame or photoionization detection. Samples as small as 0.1 cc
allow for direct stack monitoring. Because the entire air sample passes the
column and detector, ambient methane serves as an internal standard on every
analysis.
In Phase I, two PFGC instruments were constructed. These instruments were
equipped with commercial dual flame and photoionization detectors (FID/PID),
and demonstrated their ruggedness, sensitivity, and precision in monitoring
Portland’s outdoor air. Monitoring was carried out with one PFGC at an
urban monitoring station run by the Oregon Department of Environmental Quality
(DEQ). One goal was to compare the performances of the PID and the FID in a
TO-14 type environment as conducted on a typical sixth-day canister analysis
schedule. The second PFGC, used for laboratory testing and calibration, delivered
1 ppbV precision (standard error in the mean) for sub-ppbV levels of chlorobenzene.
In Phase II, VOCTEC will carry out a year-long ambient TO-14 inter-comparison
under contract to the DEQ at no cost to the Small Business Innovation Research
(SBIR) Program. This will involve DEQ’s traditional U.S. Environmental
Protection Agency-approved sixth-day TO-14 canister analysis and VOCTEC’s
24/7 in situ ambient air analysis. Additionally, the TO-14 canisters will be
analyzed using PFGC for comparison with the in situ data. In past work, VOCTEC
has followed diurnal variation of a major VOC component that was not present
when the canister was analyzed. Apparently, this compound either adsorbs or
decomposes in the canister. During Phase II, major improvements will be made
in chromatographic resolution to better separate HAPs from less toxic VOCs.
This will involve novel and proprietary software and hardware modifications
of the existing PFGC instrument.
Technology keywords
Not reported
Awards with the same topic code
Verified by an exact topic code match. This is not a text-similarity match.
- Multi-Metals Monitoring System Based Upon Spark-Induced Breakdown Spectroscopy [7215-140]
- Detecting Metals in Ambient Particulate Matter: X-Ray Fluorescence Analysis of High Volume Impaction Deposits
- Micro-Discharge Based Multi-Metal Emissions Monitoring System
- Microchip Capillary Electrophoresis for On-Line Monitoring of Inorganic Aerosols