SBIR/STTR Research Award
THROMBORESISTANT POLYURETHANE CARDIAC VALVE DEVELOPMENT
Award amount
$50,000Program and phase
SBIR · Phase IAward year
1988Award details
- Awarding agency
- Department of Health and Human Services
- Branch
- Not reported
- Recipient company
- Thoratec Laboratories Corp
- Proposal award date
- Not reported
- Contract end date
- Not reported
- UEI
- Not reported
- Solicitation number
- Not reported
- Topic code
- Not reported
Research abstract
A NEW CARDIAC VALVE FABRICATED FROM A NOVEL POLYURETHANE MATERIAL HAS BEEN FABRICATED AND TESTED IN VITRO. THESE PRELIMINARY TESTS INDICATE THAT THIS DEVICE ID DURABLE (NO FAILURES IN TESTS LASTING AT LEAST 200 MILLION CYCLES) AND POSSESSES HEMODYNAMIC CHARACTERISTICS COMPARABLE TO CURRENTLY AVAILABLE VALVES. THE SEGMENTED POLYURETHANEUREA DEVELOPED FOR THIS USE IS UNIQUE BECAUSE ITS BULK PROPERTIES, FOR LONG-TERM DURABILITY, AND SURFACE PROPERTIES, FOR BIOCOMPATIBILITY, HAVE BEEN SEPARATED AND DEVELOPED IN TWO DISTINCT MATERIALS. BLENDING OF THE TWO COMPONENTS RESULTS IN A SINGLE MATERIAL INCORPORATED THE PROPERTIES OF EACH. IN VIVO EVALUATION OF RODS OF THIS MATERIAL AND OF VENTRICULAR ASSIST DEVICE PUMPING SACS HAS DEMONSTRATED ITS EXCELLENT PHYSICAL AND THROMBORESISTANT CHARACTERISTICS. WE PROPOSE TO FABRICATE VALVES AND TEST THEM BOTH IN VITRO AND IN INVO. STEADY-STATE AND PULSATILE CONDITIONS WILL BE USED FOR THE MOCK LOOP TESTING. VALVES WILL BE IMPLANTEDIN LFFT VENTRICLE TO AORTA SHUNTS IN CALVES, A MODEL THAT WE HAVE USED MANY TIMES TO EVALUATE OTHER DEVICES. ADDITIONAL HEMODYNAMIC TESTING WILL BE PERFORMED ON EXPLANTED VALVES, AS WILL SEM AND X-RAY ANALYSIS.
Technology keywords
Not reported