Research abstract
Project Summary / Abstract
Heart failure (HF) remains a leading cause of death and results in huge economic and patient burden. Despite
policies targeted at reducing readmission rates and improved treatment options, hospitalization rates for HF have
remained largely the same. Disproportionally greater interstitial fluid accumulation versus intravascular
accumulation plays a pivotal role in the symptomatic manifestation and detrimental outcomes of HF, including
pulmonary and peripheral edema. Adequate management of fluid overload requires early intervention in the HF
“congestion cascade” prior to the onset of symptomatic congestion. Current methods used to monitor quantitative
changes in volume status are mostly limited to the intravascular space, and experimental devices targeting the
measurement of subcutaneous interstitial pressure (SCIP) have unable to provide continuous, reliable data.
There is a need for a non-invasive, easily placed technology that provides a clear quantitative metric that gives
physicians insight into when to medically intervene earlier in the HF congestion cascade. To address this critical
unmet need, NXT developed a IFPx SCIP sensor prototype (IFPx) that measures fluid status in the interstitial
fluid compartment. Our long-term goal is to develop a wearable, non-invasive device to continuously monitor
SCIP, with a form-factor like continuous glucose monitoring devices. The objective for this Phase 1 application
is to test our NXT IFPx SCIP prototype sensor in benchtop experiments and in a large animal model of fluid-
overloaded heart failure to demonstrate sensor reliability and to demonstrate SCIP signal tracking with standard,
invasive metrics of congestion. Preliminary proof-of-principle studies demonstrated that the SCIP, as measured
by historic methods, tracks with gold standard metrics of congestion, including left ventricular end diastolic
pressure (LVEDP) and central venous pressure (CVP) in a porcine model of fluid-overloaded HF. Studies in Aim
1 are designed to improve the reliability and repeatability of the NXT IFPx SCIP prototype. We will design, test,
and iterate different pressure sensing transduction devices to improve reliability and repeatability of the IFPx
prototype, as assessed by benchtop testing. Studies in Aim 2 will mainly confirm that the NXT IFPx SCIP sensor
can transmit and store data reliably. The most promising candidate from bench tests will then be validated in
acute HF animal studies in Aim 3 with success defined as changes in SCIP correlates (Phase-adjusted
correlation coefficient r2 rt 0.7) or associates (Bland-Altman average±2SD) with LVEDP or CVP Successful
completion of these studies will initiate a Phase II application to develop wearable technology further in chronic
animal studies and to perform pilot clinical studies in patients with HF.
Technology keywords
Not reported
Awards with the same topic code
Verified by an exact topic code match. This is not a text-similarity match.
- DUCTUS ARTERIOSUS STENT TO IMPROVE CONGENITAL HEART DEFECT MORTALITY
- Head Up CPR and Aortic Occlusion to Improve Survival after Refractory Cardiac Arrest
- Free-Breathing Dynamic Cardiopulmonary Function Imaging System to use with Xenon MRI
- The Next Wave of Asthma Therapeutics
- Bearingless flux reversal motor for neonatal and pediatric extracorporeal life support applications